TitleGambro AK 200 S - Service Manual
Tags Nature Wellness Hemodialysis Dialysis
File Size4.0 MB
Total Pages315
Table of Contents
                            AK 200 S - Operator’s manual
	Before you get started - general
information
		How to use this manual
			Chapter 1; Before you get started - general information
			Chapter 2; Description - the machine and its components
			Chapter 3; Controlling the machine - general guidelines
			Chapter 4; Hemodialysis, double needle treatment
			Chapter 5; Hemodialysis, single needle treatment
			Chapter 6; Isolated UF
			Chapter 7; Profiling
			Chapter 8; Hemodiafiltration, low volume
			Chapter 9; BVS - Blood Volume Sensor
			Chapter 10; BPM - Blood Pressure Monitor
			Chapter 11; Diascan
			Chapter 12; Hygiene and maintenance
			Chapter 13; Technical data and specifications
			Chapter 14; Alarm list, Attention alarm list and Technical alarm
			Chapter 15; Major changes in operator's manual
		Definitions of expressions used in this manual
		General precautions before use
		Intended use
		Safety philosophy
		The preparation of dialysis fluid
		Inlet water requirements
		Preparation of the dialysis fluid
		UFD- Ultrafiltered Dialysis Fluid (option)
		Ultrafilters; frequency of change
		List of concentrates, accessories and disposables, for use with
		Concentrates
		Lines
		Dialyzers/Ultrafilter
		Infusion fluid for low volume HDF
		Blood pressure measurement accessories
		List of symbols
		CE-marking
		Manufacturer
			Copyright
	Description - the machine and its
components
		Blood monitor
		BM 200 - Single Pump, BM 202 - Double Pump
		Fluid monitor
		FM 202 - Acetate and Bicarbonate
		Rear of the machine - without options
		Rear of the machine - options included
		Monitor stand
		MS 200: Adjustable height: five fixed positions
		MS 202: Electrically adjustable height
	Controlling the machine - general
guidelines
		Operator's panel
			1. Hygiene group
			2. Mode group
			3. Therapy group
			4. Treatment overview group
			5. Stand-by group
			6. Display and setting group
		Hygiene group
		Mode group
		Concentrate stand-by mode
		Procedure for manual activation
		Resume fluid preparation
		Therapy group
		Treatment overview group
		Bargraph display
		Flow diagram
		Time display
		Parameter buttons and their displays
		Reset possibilities
		Alarms
		Alarm indicators
			Air detector alarm
			BVS
			Attention alarm
			Technical alarm
			Blood leak alarm
			pH alarm (option)
		Alarm function of the parameter buttons
			Venous pressure alarm
			Centralizing alarm
			Blood flow rate alarm (single needle/double pump mode only)
			Temperature alarm
			Conductivity alarm
			Arterial pressure alarm
			Centralizing alarm
			Transmembrane pressure alarm
			Centralizing alarm
			UF volume alarm
			Time attention alarm
			BPM attention alarm
			Diascan attention alarm
		Override possibilities
			Arterial pressure alarm override
			Low venous pressure alarm override
			Air detector alarm override
			Blood leak alarm override
		Stand-by group
		Display and setting group
		Information display
		Treatment overview screen
		Parameters shown on Treatment overview screen
		Function check overview screen
		Tests shown on Function check overview screen
		The eight commandments
		Setting of parameters
			Venous pressure
			Blood flow in single pump mode
			Blood flow in single needle/double pump mode
			Blood pump segment adjustment
			SN data in single needle/single pump mode
			SN data in single needle/double pump
			Temperature
			Conductivity in acetate mode
			Conductivity in bicarbonate mode
			Arterial pressure
			TMP
			UF volume
			UF rate
			Time
			Heparin
			Dialysis fluid flow
			BPM (option), BVS and Diascan
		Downloading treatment parameters
		Procedure
			Accept
			Reject
		The ultrafiltration control
		Protective system for ultrafiltration
		Power failure
		Returning the blood during a power failure
		Battery back-up
		Recovery from a short power failure
		Recirculation during treatment
		Reconnection of the patient
		Change of blood pump segment
		BiCart/SelectCart change during treatment
	Hemodialysis, double needle treatment
		Double needle treatment
		Check before starting the machine
		Start the machine
		Function check
		Select concentrates
		Connect concentrates
			SelectCart and SelectBag together with BiCart
			Acidic concentrate and BiCart
			Acidic concentrate and liquid bicarbonate concentrate
			Acetate concentrate
		Start the machine - continued
		Attaching the dialyzer and the blood lines
		Dialyzer
		Arterial blood line
			Preparations if the heparin pump is to be used
		Venous blood line
		Priming
		Priming and rinsing the extracorporeal circuit
		Before priming
		Priming and rinsing procedure
		Initiating
		Set the treatment parameters
		Always check/adjust
		Connect the patient
		Start the treatment
		Pressure alarm limits setting
			Venous/Arterial pressure
			TMP
			Low venous pressure alarm limit setting below +10 mmHg
			Operating and setting
		Discontinuing
		Automatic emptying of BiCart, SelectCart and SelectBag
		Discontinuing procedure
			Confirm treatment end
			If desired, start emptying BiCart or BiCart, SelectCart and Sele
			Return the blood
			After-treatment handling of the machine
	Hemodialysis, single needle treatment
		Single needle treatment
		General
		Glossary of single needle parameters and key terms
			Single needle, double pump
			Single needle, single pump
		Single needle treatment with two blood pumps - BM 202 only
		Preparations
		Arterial blood line
		Venous blood line
		Priming
		Priming and rinsing the extracorporeal circuit
		Before priming
		Priming and rinsing procedure
			Priming and rinsing procedure in double needle mode, finishing i
			Priming and rinsing procedure in single needle/double pump mode
		Initiating
		Set the treatment parameters
		Always check/adjust
		Connect the patient
		Start the treatment
		Pressure alarm limits setting
			TMP
			Venous/Arterial pressure
			Low venous pressure alarm limit setting below +10 mmHg
			Operating and setting
		Discontinuing
			Discontinuing in double needle mode after single needle/double p
			Discontinuing in single needle/double pump mode
		Single needle treatment with one blood pump
		Preparations
		Single needle / single pump settings
		Connect the patient
			For both pressure- and time-controlled phases
		Start the treatment
			For both pressure- and time-controlled phases
			Only for pressure-controlled phases
			Only for time-controlled phases
		Discontinuing
			Discontinuing in double needle mode after single needle/single p
			Discontinuing in single needle/single pump mode
		Double needle to single needle treatment - change during treatme
		Resume treatment
		Single needle / single pump settings
			Only for pressure-controlled phases
			Only for time-controlled phases
		Discontinuing
	Isolated UF
	Profiling
		Profiling of Na + and HCO 3 - dialysis fluid concentration
		Profiling of ultrafiltration rate
		Operation
			Curve Setting/Change/Cancellation:
			Activation
			Verification
		Activation of profiling with already preset models
			Verification
	Hemodiafiltration, low volume
		General
		Protective system for low volume HDF
		The following conditions have to be met in order to fulfil the a
		Start up
		Priming and rinsing
		Initiating and starting
		Bolus infusion possibility
		Hemodiafiltration, low volume, single needle, single pump
	BVS - Blood Volume Sensor
		General
		ΔBV text information
		Alarm
		Graphs
		BVS ON/OFF
		BVS Off
		BVS On
	BPM - Blood Pressure Monitor (option)
		General
		Patient factors which affect readings
		Handling
		The BPM screens
		BPM button pressed once
		BPM button pressed twice
		Alarm
		Alarm limits setting
		Mode
		Start/Stop
		Start
		Stop
	Diascan
		General
		Application
		How the measuring process is performed by the machine
		Glossary of Diascan parameters and key terms
			Parameters:
			Key terms:
		Precautions before use
		Handling
		Screens
		The first screen
		Measurement mode
		Alarm limit for clearance
		Measurement result values
		The second screen
		Distribution volume
			Distribution volume calculation using the Watson formula
		Forecast
		Target
		Alarm limit for Kt/V
		Measurement result values when distribution volume has not been
		Second screen symbol
		Measurement result values when distribution volume has been set
		Second screen symbol
		The third screen
		Latest measured values
		Accumulated and forecast values
		Time to target
		Third screen symbols
		Alarms
		Clearance alarm limit
		Kt/V alarm limit
		Step-by-step instructions for measuring clearance and Kt/V
		Single clearance measurement check
		Procedure
		Result
		Continuous measuring of clearance
		Procedure
		Result
		Single Kt/V and clearance measurement check
		Procedure
		Result
		Continuous measuring of Kt/V and clearance
		Procedure
		Result
	Hygiene and maintenance
		General hygiene and maintenance
		Glossary of disinfection terms
		Disinfection history
		List of disinfection screen symbols
		Schedule for hygiene and maintenance
		When the machine is planned not to be in use
		Characteristics of different disinfection, decalcification and c
		Hygiene - Heat disinfection
		General - Heat disinfection
		Performing Heat disinfection program
		Before starting Heat disinfection program
		Procedure
		Performing Heat disinfection program with CleanCart
		Before starting Heat disinfection program with CleanCart
		Procedure
		Performing Heat disinfection program with liquid citric acid
		Before starting Heat disinfection program with liquid citric aci
		Procedure
		Performing integrated Heat disinfection program with WRO 300 H
		Before starting integrated Heat disinfection program with WRO 30
		Procedure
		Automatic start of Heat disinfection program
		Before automatic start of Heat disinfection program
		Procedure
		In combination with CleanCart
		Automatic switch off after Heat disinfection program
		Integrated Heat disinfection
		Hygiene - Chemical disinfection
		General - Chemical disinfection
		Chemical disinfectants
			Solutions based on peracetic acid:
			Sodium hypochlorite:
			Other disinfectants based on sodium hypochlorite:
			Formaldehyde:
		Performing Chemical disinfection program
		Before starting Chemical disinfection program
		Procedure
		Filling the machine with chemical disinfectant
		Residual test after Chemical disinfection
		Central Chemical disinfection
		Performing central Chemical disinfection
		Before starting central Chemical disinfection program
		Procedure
		Hygiene - Rinse/Drain
		Performing Rinse/Drain program
		Before starting Rinse/Drain program
		Procedure
		Performing Rinse/Drain program with CleanCart
		Before starting Rinse/Drain program with CleanCart
		Procedure
		Automatic start of Rinse/Drain program
		Before automatic start of Rinse/Drain program
		Procedure
		In combination with CleanCart
		Automatic switch off after Rinse/Drain program
		Maintenance - the flow path
		The flow path
		Cleaning
		Decalcification
		Ultrafilter; how to change
		Ultrafilter replacement reminder
		Maintenance - the exterior of the machine
		The outside
		Top tray
		Blood pump
		Blood leak detector
		BPM cuff
	Technical data and specifications
		Performance and specification - Control System
		Blood Flow Control
			Double Needle
			Single Needle Single Pump
			Single Needle Double Pump (HD only)
		Air Detection
		Heparin Administration
		Blood Volume Sensor (BVS)
		Blood Pressure Monitor (BPM)
		Dialysis fluid preparation
			Temperature
			Fluid flow
			Conductivity
			Concentration
			Concentrates
			pH supervision
			Degassing
		Dialysis fluid preparation (BiCart Select mode only)
			Fluid flow
		Ultrafiltration control
		Substitution fluid (Low volume HDF)
		Profiling
			Half-times:
		Diascan
		Disinfection and Cleaning
			Heat disinfection Program
			Heat disinfection Program with CleanCart
			Heat disinfection Program with liquid citric acid
			Heat disinfection Program with WRO 300 H
			Peracetic Acid Program
			Low concentration Peracetic Acid Program
			Hypochlorite Program
			Formaldehyde Program
			Central Disinfection Program
			Other programs, total time
		Water supply
		Drain
		Power supply
		Connection of external equipment
			External serial interface:
			RS-232
			RS-422
			External interface:
			External I/O Interface
			External interface:
		Battery Back-up
		Performance and specification - Supervisory system
		Blood Pressure Supervision
			Venous Pressure
			Arterial Pressure
			System Pressure
		Air Detection
		Heparin Administration
		Dialysis fluid preparation
			Temperature
			Conductivity
		Dialysis fluid preparation (BiCart Select mode only)
		Ultrafiltration
		Substitution fluid (Low volume HDF)
		Blood leak detection
		Physical data
		References
		Materials in contact with water, concentrates and dialysis fluid
		Polymers
		Metals
		Others
		Environmental data
		Operation
		Transportation and storage
		Electromagnetic environment
		Standards
	Alarm list, Attention alarm list and
Technical alarm list
		Alarm list
		Attention alarm list
		Technical alarm list
	Major changes in operator’s manual
		Major changes between program version 7.xx and 8.xx
		Chapter 1
		Chapter 2
		Chapter 3
		Chapter 4
		Chapter 5
		Chapter 11
		Chapter 12
		Chapter 14
	Index
                        
Document Text Contents
Page 2

Contents

1. Before you get started - general information
How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:2
Definitions of expressions used in this manual . . . . . . . . . . . . . . . . . . 1:4

General precautions before use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:5
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:8
Safety philosophy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:9
The preparation of dialysis fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:12
Inlet water requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:12
Preparation of the dialysis fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:12
UFD- Ultrafiltered Dialysis Fluid (option) . . . . . . . . . . . . . . . . . . . . . . 1:12
Ultrafilters; frequency of change . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:12

List of concentrates, accessories and disposables, for use with the
AK 200 S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:13

Concentrates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:13
Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:14
Dialyzers/Ultrafilter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:14
Infusion fluid for low volume HDF . . . . . . . . . . . . . . . . . . . . . . . . 1:15
Blood pressure measurement accessories . . . . . . . . . . . . . . . . . . . . 1:15

List of symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:16
CE-marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:18
Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:18

Copyright . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1:18

2. Description - the machine and its components
Blood monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2:2
BM 200 - Single Pump, BM 202 - Double Pump . . . . . . . . . . . . . . . . . 2:2

Fluid monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2:4
FM 202 - Acetate and Bicarbonate . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2:4

Rear of the machine - without options . . . . . . . . . . . . . . . . . . . . . . . . . . . 2:6
Rear of the machine - options included . . . . . . . . . . . . . . . . . . . . . . . . . . . 2:8
Monitor stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2:10
MS 200: Adjustable height: five fixed positions . . . . . . . . . . . . . . . . . . 2:10
MS 202: Electrically adjustable height . . . . . . . . . . . . . . . . . . . . . . . . . 2:12

3. Controlling the machine - general guidelines
Operator’s panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:2
Hygiene group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:4
Mode group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:5
Concentrate stand-by mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:6
Procedure for manual activation . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:7
Resume fluid preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:7

Therapy group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:9
Treatment overview group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:10

Bargraph display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:10
Flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:10
Time display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:11

Parameter buttons and their displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:12
Reset possibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:16

Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:17
Alarm indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3:17

HCEN9751 Revision.12.2005
Program version 8.xx

Page 314

L
Latch for filter change . . . . . . . . . . . 2:9
Level adjustment buttons . . . . . . . . 2:3
Linear graph . . . . . . . . . . . . . . . . . . 7:2
Lines . . . . . . . . . . . . . . . . . . . . . . 1:14
arterial blood line . . . . . . . . 4:12, 5:4
venous blood line . . . . . . . . 4:16, 5:5

Low Flow Heat . . . . . . . . . 12:14, 14:33

M
Machine
blood monitor . . . . . . . . . . . . . . . 2:2
BM 200 and BM 202 . . . . . . . . 2:2

fluid monitor . . . . . . . . . . . . . . . . 2:4
FM 202 . . . . . . . . . . . . . . . . . . 2:4

physical data . . . . . . . . . . . . . . 13:16
power supply . . . . . . . . . . . . . 13:10
rear of the machine
options included . . . . . . . . . . . 2:8
without options . . . . . . . . . . . . 2:6

transportation and storage . . . . 13:18
Main switch . . . . . . . . . . . . . . . 2:7, 2:9
Mains supply inlet . . . . . . . . . . . 2:7, 2:9
Maintenance
blood leak detector . . . . . . . . . 12:33
blood pump . . . . . . . . . . . . . . 12:33
BPM cuff . . . . . . . . . . . . . . . . 12:34
exterior . . . . . . . . . . . . . . . . . . 12:33
top tray . . . . . . . . . . . . . . . . . . 12:33

Manufacturer . . . . . . . . . . . . . . . . 1:18
Monitor stand
bars, lifting positions . . . . . 2:10, 2:12
height control . . . . . . . . . . . . . . 2:13
power supply cord . . . . . . . . . . . 2:13
rear handle . . . . . . . . . . . . 2:10, 2:12
screws . . . . . . . . . . . . . . . 2:11– 2:12

Mute button . . . . . . . . . . . . . . . . . 3:22

N
Note . . . . . . . . . . . . . . . . . . . . . . . . 1:4

O
ON/OFF button . . . . . . . . . . . . . . 3:4
Operator’s panel . . . . . . . . . . . . 2:3, 3:3
Optional connection . . . . . . . . . 2:7, 2:9
Override . . . . . . . . . . . . . . . . . . . . 3:20
air detector alarm . . . . . . . . . . . 3:20
arterial pressure alarm . . . . . . . . 3:20
blood leak alarm . . . . . . . . . . . . 3:21
low venous pressure alarm . . . . . 3:20

P
Parameter buttons . . . . . . . . . . . . . 3:12
Parameter settings

arterial pressure . . . . . . . . . . . . . 3:31
blood flow . . . . . . . . . . . . . . . . 3:30
blood pump segment . . . . . . . . . 3:30
BPM, BVS, Diascan . . . . . . . . . 3:31
conductivity . . . . . . . . . . 3:30– 3:31
dialysis fluid flow . . . . . . . . . . . . 3:31
heparin . . . . . . . . . . . . . . . . . . . 3:31
SN data . . . . . . . . . . . . . . . . . . 3:30
temperature . . . . . . . . . . . . . . . 3:30
time . . . . . . . . . . . . . . . . . . . . . 3:31
TMP . . . . . . . . . . . . . . . . . . . . . 3:31
UF rate . . . . . . . . . . . . . . . . . . . 3:31
UF volume . . . . . . . . . . . . . . . . 3:31
venous pressure . . . . . . . . . . . . 3:30

Per-Low . . . . . . . . . . . . . . . . . . . 12:20
Peracetic acid . . . . . . . . . . . . . . . 12:19
Physical data of the machine . . . . 13:16
Pick-up tube
A pick-up tube . . . . . . . . . . . . . . 2:4
B pick-up tube . . . . . . . . . . . . . . 2:4

Potential equalization . . . . . . . . . . . 1:6
Potential equalization connection . . 2:6,
2:8
Power failure
recovery . . . . . . . . . . . . . . . . . . 3:39
returning the blood . . . . . . . . . . 3:38

Power supply unit . . . . . . . . . . . 2:7, 2:9
Precautions . . . . . . . . . . . . . . . . . . . 1:5
Preparation of dialysis fluid . . . . . . 1:12
Preset
can be preset . . . . . . . . . . . . . . . . 1:4

Priming
in double needle mode, finishing
in single needle/double pump
mode . . . . . . . . . . . . . . . . . . . . 5:6
in HDF low volume . . . . . . . . . . 8:6
in single needle/double pump
mode . . . . . . . . . . . . . . . . . . . . 5:9
priming button . . . . . . . . . . . . . . 3:9
priming detector . . . . . . . . . . . . . 2:3

Profiling
activation . . . . . . . . . . . . . . . . . . 7:7
degressive graph . . . . . . . . . . . . . 7:2
half-time . . . . . . . . . . . . . . . . . . . 7:2
linear graph . . . . . . . . . . . . . . . . . 7:2
profiling button . . . . . . . . . . . . . . 3:9
progressive graph . . . . . . . . . . . . 7:2
technical data . . . . . . . . . . . . . . 13:5
UF rate . . . . . . . . . . . . . . . . . . . . 7:4

Progressive graph . . . . . . . . . . . . . . 7:2
Protective system . . . . . . . . . . . . . . 1:9
for low volume HDF . . . . . . . . . . 8:2
for ultrafiltration . . . . . . . . . . . . 3:35

Pulse rate symbol . . . . . . . . 10:6– 10:7

HCEN9751 Revision.12.2005
Program version 8.xxAK 200 S Operator’s manual - Index

Page 315

R
Rear of the machine
options included . . . . . . . . . . . . . 2:8
without options . . . . . . . . . . . . . . 2:6

Recirculation . . . . . . . . . . . . . . . . . 3:40
Relay interface connection . . . . . 2:6, 2:8
Reset . . . . . . . . . . . . . . . . . . . . . . 3:16
Residual test . . . . . . . . . . . . . . . . 12:23
Rinse/Drain button . . . . . . . . . . . . 3:4
Rinse/Drain program . . . . . . . . . 12:25
automatic start . . . . . . . . . . . . 12:28
automatic switch off . . . . . . . . 12:29
with CleanCart . . 12:17, 12:26, 12:29

S
Safety coupling . . . . . . . . . . . . . . . . 2:4
Safety philosophy . . . . . . . . . . . . . . 1:9
SelectBag
automatic emptying . . . . . . . . . . 4:24
holder . . . . . . . . . . . . . . . . . . . . . 2:5
how to connect . . . . . . . . . . . . . . 4:7

SelectCart
automatic emptying . . . . . . . . . . 4:24
change during treatment . . . . . . 3:43
holder . . . . . . . . . . . . . . . . . . . . . 2:5
how to connect . . . . . . . . . . . . . . 4:7

Set knob . . . . . . . . . . . . . . . . . . . . 3:23
Single needle
parameters and key terms . . . . . . 5:3

Single needle button . . . . . . . . . . . . 3:5
Smiley symbol . . . . . . . . . . . . . . . 11:14
Sodium hypochlorite . . . . . . . . . . 12:19
Standards . . . . . . . . . . . . . . . . . . 13:22
Start UF Stop button . . . . . . . . . . . 3:22
Substitution fluid . . . . . . . . 13:5, 13:15
Symbols
arrows symbol . . . . . . . . . . . . . 11:14
clock symbol . . . . . . . . . . 10:6, 11:14
goal flag symbol . . . . . .11:11– 11:12
list of symbols . . . . . . . . . . . . . . 1:16
pulse rate symbol . . . . . . 10:6– 10:7
screen symbols, disinfection . . . . 12:4
smiley symbol . . . . . . . . . . . . . 11:14
time to target symbol . . . . . . . . 11:14

System pressure transducer
connector . . . . . . . . . . . . . . . . 2:3, 5:5
System pressure transducer protector 8:4

T
Time display . . . . . . . . . . . . . . . . . 3:11
Time to target symbol . . . . . . . . . 11:14
Top tray . . . . . . . . . . . . . . . . . . . . . 2:2
how to maintain . . . . . . . . . . . 12:33

Tray for concentrate containers . . .2:11,
2:13
Treatment overview screen . . . . . . 3:24
Treatment parameters
download . . . . . . . . . . . . . . . . . 3:32

U
UF calibration unit connection . 2:7, 2:9
UF rate profiling . . . . . . . . . . . . . . . 7:4
Ultrafiltered Dialysis Fluid . . . . . . . 1:12
Ultrafilters
frequency of change . . . . . . . . . 1:12
ultrafilter . . . . . . . . . . . . . . . . . . 1:14
how to change . . . . . . . . . . . 12:31

Ultrafiltration
control . . . . . . . . . . . . . . . . . . . 3:34
control, technical data . . . . . . . . 13:5
technical data . . . . . . . . . . . . . 13:15
ultrafiltration path . . . . . . . . . . . 3:10

V
Venous blood line . . . . . . . . . . 4:16, 5:5
Venous blood pump . . . . . . . . . 2:3, 5:5
Venous expansion chamber . . . . . . . 5:5
Venous line clamp . . . . . . . . . . 2:3, 4:16
Venous pressure transducer
connector . . . . . . . . . . . . . . . 2:2, 4:16
Volume control button . . . . . . . . . . 3:5

W
Warning . . . . . . . . . . . . . . . . . . . . . 1:4
Water inlet . . . . . . . . . . . . . . . . 2:7, 2:9
Watson formula . . . . . . . . . . . . . . 11:9
Wheels, lockable . . . . . . . . . . 2:11, 2:13

HCEN9751 Revision.12.2005
Program version 8.xx AK 200 S Operator’s manual - Index